271
Users Manual
11-4
offset changes determines a practical minimum which is greater than this. Recommended
times are at least 10 ms for frequency plus wave shape changes and at least 50 ms for
amplitude and offset changes.
If a shorter duration than that recommended above is set the results will be unpredictable
and it is likely that hop cannot be turned off in the usual way. To recover from this
situation hold the
ESCAPE
key down for approximately one second until hop mode is
exited.
Saving Hop Settings
The current hop setting is saved in non-volatile memory at power-down. It is not part of
the data saved by the
STORE
function (see
Storing and recalling set-ups
in chapter 12,
System Operations
) and therefore only one complete hop sequence can be stored.
The hop setting is maintained when the system defaults are reloaded.
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for DDS Function Generator 271
Page 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...