271
Users Manual
1-6
Trigger and burst
Phase-coherent signal keying: each positive edge of the trigger signal will produce one
burst of the carrier, starting and stopping at the phase angle specified by the start/stop
phase setting.
Carrier frequency:
0.1 mHz to >1 MHz
Carrier waveforms:
All
Number of cycles:
1 to 1023 (resolution 1 cycle)
or 0.5 to 511.5 (resolution 0.5 cycle).
Trigger repetition rate:
dc to 50 kHz internal, dc to 1 MHz external.
Source:
Manual (front panel key), internal trigger generator, external
signal or remote interface.
Gated
Non-phase coherent signal keying: the output carrier wave is on while the gate signal is
high and off while it is low.
Carrier frequency:
From 0.1 mHz to 10 MHz.
Carrier waveforms:
All
Trigger repetition rate:
dc to 50 kHz internal, dc to 1 MHz external.
Gate signal source:
Manual (front panel key), internal trigger generator, external
signal or remote interface.
Sweep
Carrier waveforms:
All
Sweep modes:
Linear or logarithmic, single or continuous.
Sweep width:
From 0.1 mHz to 10 MHz in one range. Phase continuous.
Start and stop frequency may be set independently.
Sweep time:
10 ms to 999 s with 3 digit resolution.
Markers:
Two, variable during sweep, available at the rear panel
socket.
Sweep trigger source:
The sweep may be free run or triggered manually (front
panel key), by an external signal or through a remote
interface.
Amplitude Modulation
Carrier frequency:
From 0.1 mHz to 10 MHz.
Carrier waveforms:
All.
Depth:
Variable 0 to 100% typical, resolution 1 %.
Internal source:
1 kHz fixed sine wave or 0.005 Hz to 50 kHz square wave.
External: See
VCA In
below.
Frequency Shift Keying (FSK)
Phase coherent switching between two selected frequencies at a rate defined by the
switching signal source.
Carrier frequency:
From 0.1 mHz to 10 MHz.
Carrier waveforms:
All.
Switch repetition rate:
dc to 50 kHz internal, dc to 1 MHz external.
Switching signal source:
Manual (front panel key), internal trigger generator, external
signal or remote interface.
1.888.475.5235
Fluke-Direct
.com
Summary of Contents for DDS Function Generator 271
Page 12: ...271 Users Manual x 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 22: ...271 Users Manual 1 10 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 34: ...271 Users Manual 4 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 46: ...271 Users Manual 5 12 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 52: ...271 Users Manual 6 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 60: ...271 Users Manual 8 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 64: ...271 Users Manual 9 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 76: ...271 Users Manual 12 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 80: ...271 Users Manual 13 4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 90: ...271 Users Manual 15 6 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 110: ...271 Users Manual 16 20 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 112: ...271 Users Manual 2 1 888 475 5235 info Fluke Direct com Fluke Direct com ...
Page 132: ...4 1 888 475 5235 info Fluke Direct com Fluke Direct com ...