Comark RF500A/AP Service Manual
Tier 2: RF500A Module Replacements
Replace the Battery Pack
Completely disassemble the RF500A/AP as per the Disassembly section.
NOTE: Self adhesive pads should be cut with sharp knife when lifting the battery pack.
The battery pack is held in place by 2 cable ties that pass through 4 holes in the PCB. Additionally there are 4 double
sided adhesive foam pads under the pack.
To remove the pack, cut the cable ties:
Finally carefully but firmly lever off the battery pack
17 November 2011