Comark RF500A/AP Service Manual
Tier 3: RF500A Component Level Repair
Additional Fault Diagnostic Information
The RF500A/AP web interface can provide additional information for intermittent type faults which can be more
troublesome to identify.
Hardware Status Page
Only the “
System
” section of the hardware page will be discussed here, for further information regarding the rest of the
page consult the web view manual.
ID
Item
Comments
PC
5V Rail
The rail voltage as measured by the ADC on the MPU for more
information regarding the rail see Main 5V Rail
micro
3VC Rail
The rail voltage as measured by the ADC on the MPU for more
information regarding the rail see Permanent 3.3V Rail
pwr line
Common Power Rail
The rail voltage as measured by the ADC on the MPU for more
information regarding the rail see Power Supply
case
Case Temperature
The temperature as measured by the RTC, used to monitor the
internal case temperature for more information see Real Time Clock
(RTC)
por
Power on Reset Event
This is the number of times the MPU has been reset due to all power
(including battery pack) being completely removed and restored.
Once this has left production this should be always be 0, anything
else would indicate a serious fault or tempering.
wtd
Watchdog Reset Event
This is the number of times the MPU has been reset due to a
watchdog reset, this should generally be 0 but a count of less than 5 is
acceptable.
jtag
Jtag Reset Event
This is the number of times the MPU has been reset due to a Jtag
reset event, this should always be 0.