Comark RF500A/AP Service Manual
Appendix C – Website Error Messages
Appendix C – Website Error Messages
No Hardware Pass message
Actual Message:
No Hardware Pass message
The RF500A/AP has detected that the control PCBA has not been tested correctly at the manufacturing stage, the board
must be returned to the PCBA manufacturer and run through the Comark PCBA test equipment.
SD cards appear to be in the wrong slots...
Actual Message:
SD cards appear to be in the wrong slots,
shutdown and swap SD cards
The RF500A/AP has detected that SD1 and SD2 have been swapped, shutdown the unit and swap the cards and reboot.
SD cards appear to have different data ID tags
Actual Message:
SD cards appear to have different data ID tags
The RF500A/AP has detected that the fitted SD cards are not a matched pair (hence have different ID tags), find the
appropriate matching SD card or, if there is no data involved, replace with two new blank SD cards and reboot.
Looks like you have no data on the master SD....
Actual Message:
Looks like you have no data on the master SD card but have data on the backup card.
If you are sure this is correct and want to restore data from backup SD card type
http://your-ip-address/cgi-bin/sdbackuprestore
This is the message when replacing the master SD card (SD2) as described in the Replace μSDHC Card 2 section.
booted up from backup SD card...
Actual Message:
booted up from backup SD card,
check SD card slots possible error with SD2
The RF500A/AP has booted up from the backup SD card (SD1), this indicates there is a problem the master SD
card(SD2). Replace the master SD card as described in the Replace μSDHC Card 2 section.