Comark RF500A/AP Service Manual
Tier 2: RF500A Module Replacements
4. Run Tftpd32 in the background as this will show transfers and the test PC ip address needed later.
5. From the start menu go to the Comark program group and run “RF500A Driver Update Util”
6. Enter required data by driver update util:
7. Click “Update MXM6410 Driver 1 and
follow the on screen prompts carefully
(the module reset button is
S501 in the top corner to the right of the RF PCB), when the update is complete the software will display the
following and will then ask you to reset the module:
8. Click “Update MXM6410 Driver 2 and
follow the on screen prompts carefully
, when the update is complete
the software will display the following and will then ask you to reset the module:
9. The DUT can now be powered down:
•
Remove USB Cable
•
Remove Ethernet Cable
•
Remove External Power
•
Finally press and hold Module Reset Button (S501) whilst removing battery pack plug
•
Wait 1 second
•
Release Module Reset Button (S501)
•
Remove Watchdog disable header
•
Plug in Battery Pack
17 November 2011
Page 34 of 65
Serial number of DUT
IP Address to set the DUT to
during this process
IP Address of test PC
obtained from Tftpd32 above