Comark RF500A/AP Service Manual
Appendix C – Website Error Messages
check SD card slots possible error with SD2...
Actual Message:
check SD card slots possible error with SD2 or card in wrong slot
The RF500A/AP has detected the master SD card but cannot access it. Replace the master SD card as described in the
Replace μSDHC Card 2 section.
check SD card slots possible error with SD1...
Actual Message:
check SD card slots possible error with SD1 or card in wrong slot
The RF500A/AP has detected the slave SD card but cannot access it. Replace the slave SD card as described in the
Replace μSDHC Card 1 section.
File Error
There are various different types of file error message which will not be detailed here, most are indicative of memory
corruption meaning the safest option is to replace the memory cards and restore any data from a backup as described in
Appendix D – Restore Data to an RF500A/AP.