![Fluke CableAnalyzer DSP-4000 Series User Manual Download Page 74](http://html1.mh-extra.com/html/fluke/cableanalyzer-dsp-4000-series/cableanalyzer-dsp-4000-series_user-manual_2309835074.webp)
DSP-4000 Series
Users Manual
3-14
Pressing
#
View Plot
produces the NEXT plot screen. Figure 3-5
describes an example of the screen.
6
5
4
3
2
1
oy11c.eps
1
The cable pairs relevant to the plot.
2
Frequency range in MHz. Use
U D
to switch between scales. To switch to the next lowest
or highest range, use
L R
to move the cursor beyond the left or right side of the plot.
3
The NEXT level, frequency, and margin at the cursor’s position. The cursor aligns to the
frequency that produced the worst margin. Margin is the difference between the limit and
measured values plotted at the cursor’s position. If you move the cursor beyond the highest
test frequency specified by the selected test standard, the readout shows the NEXT value
at the cursor’s position.
4
The limits for NEXT, as defined by the selected test standard. A crosshair is shown if the
limit is defined for only one frequency.
5
Decibels of crosstalk attenuation (NEXT) between the cable pairs.
6
The measured NEXT for the cable pairs.
Figure 3-5. The NEXT Plot Screen
Summary of Contents for CableAnalyzer DSP-4000 Series
Page 6: ...DSP 4000 Series Users Manual iv...
Page 8: ...DSP 4000 Series Users Manual vi...
Page 18: ...DSP 4000 Series Users Manual 1 8...
Page 60: ...DSP 4000 Series Users Manual 2 42...
Page 128: ...DSP 4000 Series Users Manual 6 6...
Page 160: ...DSP 4000 Series Users Manual 7 32...
Page 186: ...DSP 4000 Series Users Manual 8 26...
Page 188: ...DSP 4000 Series Users Manual...
Page 204: ...DSP 4000 Series Users Manual B 14 oy81f bmp Figure B 5 Autotest Report in Graphical Format...
Page 210: ...DSP 4000 Series Users Manual B 20...
Page 222: ...DSP 4000 Series Users Manual C 12...