
9 Measurement
FLUXUS F808
9.2 Measurement settings
95
UMFLUXUS_F808V2-2EN, 2020-06-25
Terminal assignment
The terminals for the connection of the output are displayed.
• Press ENTER.
If the transmitter possesses a switchable current output, it is displayed whether it is active or passive (here: active).
Output function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Enter a test value. It has to be within the output range.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
9.2.4
Activation of an analog output
The measuring range of the source item has to be entered.
• Press ENTER until
Current Loop
is displayed. Select
yes
to activate the output.
• Press ENTER.
Measuring range
After an analog output has been activated in the program branch
Output Options
, the measuring range of the source
item has to be entered.
• Select
sign
if the sign of the measured values is to be considered for the output.
• Select
absolut
if the sign of the measured values is not to be considered for the output.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. outputs\...\Active loop I1
Special Funct.\SYSTEM settings\Proc. outputs\...\Output Test I1
Special Funct.\SYSTEM settings\Proc. outputs\...\I1= 10 mA\Again?
Output Options
ENTER until
Current Loop
Meas.Values
Zero-Scale Val.
Full-Scale Val.
Test output
Error-val. delay
Notice!
An output can only be activated in the program branch
Output Options
if it has previously been installed.
Output Options\...\Current Loop
Output Options\...\Meas.Values