15 Data transmission
15.1 FluxDiagReader/FluxDiag
FLUXUS F808
2020-06-25, UMFLUXUS_F808V2-2EN
122
15
Data transmission
The data can be transmitted to the PC via the communication interface (process interface RS485 (optional) or service in
-
terface RS232) of the transmitter.
15.1
FluxDiagReader/FluxDiag
By means of the program FluxDiagReader it is possible to transmit transmitter data via service interfaces to the PC. The
transmission takes place via the RS232 service interface. For this, the measurement has to be stopped.
By means of the program FluxDiagReader it is possible to display and to store measurement data, setup settings and
snaps.
For the operation of the program see FluxDiagReader support.
For the connection of the service interface see section 7.4.
15.2
Terminal program
If FluxDiag is not available, the measurement data can be transmitted to a terminal program in ASCII format.
15.2.1
Online transmission
The measured values are transmitted during the measurement.
The data logger works independently of the online transmission but with the same transmission rate.
• Start the terminal program.
• Enter the transmission parameters into the terminal program. The transmission parameters of the terminal program and
the transmitter have to be identical, see section 15.3.
Transmitter settings
• Select the program branch
Output Options
.
• Press ENTER until the menu item
Serial Output
is displayed.
• Select
yes
to activate the online transmission.
• Press ENTER.
• Enter the storage rate.
• Start the measurement.
Tab. 15.1: Data transmission overview
program
data transmission
see
FluxDiagReader
RS232
offline
section 15.1
FluxDiag (optional)
RS232
offline or online
section 15.1
terminal program
RS232 or RS485
offline or online
section 15.2
Notice!
For the setup of the measuring point, it is possible to display and to store the current measurement data via the pro
-
gram FluxDiag (optional) on the PC. A permanent data transmission via FluxDiag is not recommended.
Notice!
It is recommended to use the RS485 interface for the online transmission of data. The RS232 interface should only
be used if the transmitter does not have an RS485 interface.
Output Options\Serial Output