13 Outputs
FLUXUS F808
13.1 Installation of a binary output
115
UMFLUXUS_F808V2-2EN, 2020-06-25
Function test of the binary output
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Select
Reed-Relay OFF
or
Open collect OFF
in the scroll list
Output Test
to test the de-energized state of the
output.
• Press ENTER. Measure the resistance at the output. The value has to be high ohmic.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
• Select
Reed-Relay ON
or
Open collect. ON
in the scroll list
Output Test
to test the energized state of the output.
• Press ENTER. Measure the resistance at the output. The value has to be low ohmic.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. outputs\...\B1 Output Test\Reed-Relay OFF
Special Funct.\...\B1 Output Test\B1=ON\Again?
Special Funct.\SYSTEM settings\Proc. outputs\...\B1 Output Test\Reed-Relay ON
Special Funct.\...\B1 Output Test\B1=ON\Again?