17 SuperUser mode
17.8 Manual reset of the totalizers
FLUXUS F808
2020-06-25, UMFLUXUS_F808V2-2EN
146
Total digits = Automatic
The number of decimal places will be adjusted dynamically. Low totalizer values will initially be displayed with 3 decimal
places. If the values of the totalizers are higher, the number of decimal places will be reduced.
Total digits = Fixed to x digit
The number of decimal points is constant. The max value of the totalizer is reduced with the number of decimal places.
17.8
Manual reset of the totalizers
If the manual reset of the totalizers is activated, the totalizers can be reset to zero during the measurement by pressing
key CLR 3 times, even if a program code is activated.
• Select
Special Funct.\SYSTEM settings\Measuring\Miscellaneous
.
• Press ENTER until the menu item
3xC
clear totals
is displayed.
• Select
on
to activate the manual reset of the totalizers,
off
to deactivate it.
• Press ENTER.
max. value
display
< 10
6
±0.000
…
±999999.999
< 10
7
±1000000.00
…
±9999999.99
< 10
8
±10000000.0
…
±99999999.9
< 10
10
±1000000000
…
±9999999999
decimal places
max. value
max. display
0
< 10
10
±9999999999
1
< 10
8
±99999999.9
2
< 10
7
±9999999.99
3
< 10
6
±999999.999
4
< 10
5
±99999.9999
Notice!
The number of decimal places and the max. value defined here only affect the display of the totalizers.
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\3xC clear totals
Notice!