18 Settings
FLUXUS F808
18.1 Dialogs and menus
149
UMFLUXUS_F808V2-2EN, 2020-06-25
18.1.2
Fluid pressure
The dependence of the parameters of a fluid on the pressure can be taken into account.
• Select
on
to enter the fluid pressure in the program branch
Parameter
. Select
off
if 1 bar is to be used for all calcula
-
tions.
• Press ENTER.
18.1.3
Measuring point number
• Select (
1234
) if the measuring point is to be identified only by numbers, point and hyphen.
• Select (
↑↓← →
) if the measuring point is to be designated with ASCII characters.
• Press ENTER.
18.1.4
Transducer distance
recommended setting:
user
•
user
will be selected if the measuring point is always the same
•
auto
can be selected if the measuring point often changes
In the program branch
Measuring
, the recommended transducer distance will be displayed in parentheses, followed by
the entered transducer distance if the recommended and the entered transducer distance are not identical.
During the transducer positioning, in the program branch
Measuring
:
• only the entered transducer distance will be displayed if
Transd. Distance
=
user
is selected and the recommended
and the entered transducer distances are identical
• only the recommended transducer distance will be displayed if
Transd. Distance
=
auto
is selected
18.1.5
Error delay
The error delay is the time interval after which the entered value for the error value is transmitted to the output in case no
valid measured values are available.
• Select
damping
if the damping factor is to be used as the error delay. Select
edit
to activate the menu item
Error-
val. delay
in the program branch
Output Options
. From now on, the error delay can be entered in the program
branch
Output Options
.
• Press ENTER.
Special Funct.\...\Dialogs/Menus\Fluid pressure
Notice!
For documentation purposes, it is useful to enter the fluid pressure, even if the transmitter contains no pressure-
dependent characteristic curves.
Special Funct.\...\Dialogs/Menus\Meas.Point No.
Special Funct.\...\Dialogs/Menus\Transd. Distance
Special Funct.\...\Dialogs/Menus\Error-val. delay
Transd. Distance
(50.8) 50.0
mm