9 Measurement
FLUXUS F808
9.2 Measurement settings
91
UMFLUXUS_F808V2-2EN, 2020-06-25
9.2.2
Input of the damping factor
Each displayed measured value is a floating average of all measured values of the last x seconds, with x being the damping
factor. A damping factor of 1 s means that the measured values are not averaged because the measuring rate is approx
1/s. The default value of 10 s is appropriate for normal flow conditions. Values which fluctuate strongly due to a higher flow
dynamic, require a higher damping factor.
• Select the program branch
Output Options
.
• Press ENTER until the menu item
Damping
is displayed.
• Enter the damping factor.
• Press ENTER.
• Press key BRK to return to the main menu.
9.2.3
Installation of an output
If the transmitter is equipped with outputs, they have to be
installed and activated before they can be used:
• assignment of the physical quantity (source item) to be
transmitted to the output by the source channel, and the
properties of the signal
• definition of the output behavior in case no valid mea
-
sured values are available
• activation of the installed output in the program branch
Output Options
• Select
Special Funct.\SYSTEM settings\Proc. outputs
.
• Press ENTER.
Selection of an output
• Select the output to be installed.
• Press ENTER.
The scroll list contains all available outputs of the transmitter:
– Current Ix (--)
– Binary Bx (--)
– Frequency Fx (--)
A tick (
) after a list item indicates that this output has already been installed.
• Select
yes
to install or reconfigure the output.
• Press ENTER.
Output Options\...\Damping
SYSTEM settings
Proc. outputs
Install Output
Enable
Source item
Notice!
The settings will be stored at the end of the dialog. If the menu item is quit by pressing key BRK, the changes will not
be stored.
Special Funct.\SYSTEM settings\Proc. outputs
Special Funct.\SYSTEM settings\Proc. outputs\Install Output
Special Funct.\SYSTEM settings\Proc. outputs\Enable I1