14 Data logger
FLUXUS F808
14.1 Activation/Deactivation of the data logger
117
UMFLUXUS_F808V2-2EN, 2020-06-25
14
Data logger
The transmitter has a data logger which stores the data during the measurement.
The following data can be stored:
• date
• time
• measuring point number
• pipe parameters
• fluid parameter
• transducer data
• sound path (reflection or diagonal arrangement)
• transducer distance
• damping factor
• storage rate
• physical quantity
• unit of measurement
• values of the totalizers (in case the totalizers are activated)
• diagnostic values (in case the storing of diagnostic values is activated)
In order to store the data, the data logger has to be activated.
The available data logger can be displayed.
14.1
Activation/Deactivation of the data logger
• Select the program branch
Output Options
.
• Press ENTER until
Store Meas.Data
is displayed.
• Select
yes
to activate the data logger,
no
to deactivate it.
• Press ENTER.
14.2
Setting the storage rate
The storage rate is the frequency to transmit or store measured values. If the storage rate is not set, the storage rate pre
-
viously selected will be used.
The recommended storage rate is 4 s.
• Select a storage rate or
EXTRA
.
• Press ENTER.
This display will only be indicated if
Store Meas.Data
and/or
Serial Output
are activated.
• Enter the storage rate if
EXTRA
was selected.
• Press ENTER.
14.3
Configuration of the data logger
• Select
Special Funct.\SYSTEM settings\Storing
.
• Press ENTER.
Output Options\...\Store Meas.Data
Output Options\...\Storage Rate
Output Options\...\Storage Rate\EXTRA
Special Funct.\SYSTEM settings\Storing