14 Data logger
FLUXUS F808
14.3 Configuration of the data logger
119
UMFLUXUS_F808V2-2EN, 2020-06-25
Totalizer storing
It is possible to store the currently displayed totalizer only or to store one value for each flow direction.
• Select
one
to store the currently displayed totalizer value only. This can apply for the positive and negative totalizer.
• Select
both
to store the totalizer values for both flow directions.
• Press ENTER.
Signal amplitude storing
• Select
on
to store the amplitude of the measured signal together with the measured values.
• Press ENTER.
Fluid sound speed storing
• Select
on
to store the fluid sound speed together with the measured values.
• Press ENTER.
Diagnostic values storing
• Select
on
to store the diagnostic values together with the measured values.
• Press ENTER.
Option of flow velocity storing
• Enter the HotCode
007043
.
• Select
auto
to store the flow velocity as an integer. Select
full
to store the flow velocity as a floating-point number.
• Press ENTER.
Notice!
Storage mode
=
average
The average of the physical quantity and other measurands assigned to the measuring channel, will be calculated.
If the storage rate < 5 s is selected,
sample
will be used.
If no average could be calculated over the complete storage interval, the value will be marked as invalid. The ASCII
file will contain
???
for invalid average values of the data and
?UNDEF
instead of invalid temperatures.
Special Funct.\SYSTEM settings\Storing\Quantity Storage
Special Funct.\SYSTEM settings\Storing\Store Amplitude
Special Funct.\SYSTEM settings\Storing\Store c-Medium
Special Funct.\SYSTEM settings\Storing\Store diagnostic
Storage resolut.
auto
>FULL<