17 SuperUser mode
FLUXUS F808
17.1 Activation/deactivation
141
UMFLUXUS_F808V2-2EN, 2020-06-25
17
SuperUser mode
The SuperUser mode offers the possibility of an advanced diagnostic of signals and measured values as well as the defi
-
nition of additional parameters adapted to the measuring point, in order to achieve better measured values or for experi
-
mental work. Special features of the SuperUser mode are:
• Default settings will not be used.
• There are no plausibility tests during the parameter entry.
• It is not checked whether the entered parameters are within the limits given by physical laws and technical data.
• The cut-off flow is not activated.
• A number of sound paths has to be entered.
Some menu items that are not visible in the normal mode are displayed.
17.1
Activation/deactivation
• Enter the HotCode
071049
, see section 8.4.
• Press ENTER.
It is displayed that the SuperUser mode is activated.
• Press ENTER. The main menu will be displayed.
• Enter the HotCode
071049
again to deactivate the SuperUser mode.
It is displayed that the SuperUser mode is deactivated.
• Press ENTER. The main menu will be displayed.
17.2
Transducer parameters
In the SuperUser mode, the menu item
Transducer Type
will be displayed in the program branch
Parameter
at the
end of the input even if the transducers are detected by the transmitter.
• Press ENTER.
or:
• Select
Special Version
to enter the transducer parameters.
• Press ENTER.
Notice!
The SuperUser mode is intended for experienced users with advanced application knowledge. The parameters can
affect the normal measuring mode and lead to wrong measured values or to a measurement failure when a new mea
-
suring point is set up.
SUPERUSER MODE\IS ACTIVE NOW
SUPERUSER MODE\IS PASSIVE NOW
Notice!
Some of the defined parameters are still activated after the deactivation of the SuperUser mode.
Parameter\...\Transducer Type\Q2E-314
Parameter\...\Transducer Type\Special Version