16 Advanced functions
FLUXUS F401
16.7 Profile correction
109
UMFLUXUS_F401V2-1EN, 2020-06-12
16.7
Profile correction
The following settings can be made for the calculation of the fluid mechanics calibration
factor k
Re
:
•
off
: profile correction 1.0
•
on
: profile correction 2.0 at ideal inflow conditions (default)
•
With disturbance
: profile correction 2.0 at non ideal inflow conditions
The following steps are necessary to set the profile correction:
• Selection of the profile correction setting for all measuring channels in the program
branch
Special Funct.
• Input of the disturbance distance in the program branch
Parameter
if
With disturbance
has been selected
If
With disturbance
has been selected, the transducers have to be mounted in
reflection arrangement, X arrangement or displaced X arrangement to compensate
transverse flow effects. When mounting in X arrangement, it is essential to set the same
parameters for both measuring channels and to activate for them a calculation channel
with average generation.
Selection of the setting
• Select the menu item
Special Funct.
in the program branch
Measuring
.
• Press ENTER until the menu item
ProfileCorr 2.0
is displayed.
• Select a list item (default:
on
).
• Press ENTER.
Input of the disturbance distance
If
With disturbance
is selected in the menu item
Special Funct.\SYSTEM
settings\Measuring\ProfileCorr 2.0
, the disturbance distance has to be
entered in the program branch
Parameter
.
• Enter the disturbance distance.
• Press ENTER.
Measurement
When starting the measurement, it is checked whether the measurement arrangement is
appropriate.
Special Funct.\...\Measuring\ProfileCorr 2.0
Disturb.distance
2.3
m