7 Connection
7.2 Power supply
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
40
Removal
• Pull the knurled ring slightly back and remove the connector from the transmitter socket.
• Put the protective cap on the connector. The red point on the connector has to be
aligned with the red marking on the protective cap.
• Put the protective cap on the transmitter socket.
7.2
Power supply
The transmitter can be operated with the integrated battery, the power supply unit or the
power pack PP026NN, see document QSPowerPack_PP026.
7.2.1
Battery operation
The transmitter has a Li-ion battery and can therefore be operated independently. When
delivered, the battery is charged with approx. 30 %. The battery does not need to be fully
charged before it is used for the first time.
If the LED of the status indicator "power supply" flashes red the battery is almost empty.
The capacity is sufficient for storing of the current parameter record. A measurement is
impossible.
Fig. 7.2:
Transducer connector
1 – protective cap with marking
2 – knurled ring
3 – marking on the transmitter socket
4 – marking on the connector
Notice!
It is recommended to discharge and subsequently recharge the batteries completely
for at least once a year.
2
4
1
pulling direction
3
Summary of Contents for FLUXUS F401
Page 1: ...FLUXUS F401 UMFLUXUS_F401V2 1EN Operating instruction...
Page 158: ...Annex D Conformity declarations FLUXUS F401 2020 06 12 UMFLUXUS_F401V2 1EN 158...
Page 159: ......
Page 160: ......