17 SuperUser mode
17.3 Setting of the measuring time in the low-power mode
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
124
17.3
Setting of the measuring time in the low-power mode
The measured values are determined during the measuring time. The current value will
be stored in the data logger the end of the measuring time.
The measuring time in the low-power mode can be defined. The default value of 5 s is
appropriate for normal flow conditions.
• Select
Special Funct.\SYSTEM settings\Measuring\Miscellaneous
.
• Press ENTER until the menu item
Measuring time
is displayed.
• Enter the measuring time. The max. measuring time is 60 s.
• Press ENTER.
This display will only be indicated if the low-power mode is activated.
17.4
Setting of measuring cycles in the night-flow mode
In the SuperUser mode it is possible to switch between a 24 h cycle, see section 16.2,
and a user-defined cycle. The user-defined cycle allows the definition of several measur
-
ing and resting phases per day.
If the night-flow mode is activated, the menu item
Daily repeat
will be displayed at the
beginning of the measurement.
• Select
yes
to enter a starting and measuring time and
no
to define the measuring and
resting phase for the night-flow mode.
• Press ENTER.
• Select
asap
to start the user-defined cycle with the next full minute and
setup
to de
-
fine the starting time.
• Press ENTER.
Special Funct.\SYSTEM settings\Measuring\Miscellaneous
Notice!
If a measuring time < 5 s is entered, the battery life of the transmitter is reduced.
Measuring\...\Nightflow mode\Daily repeat
Measuring time
5
s
Start of measure
>SETUP<
asap