10 Troubleshooting
10.6 Problems with the totalizers
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
80
10.6
Problems with the totalizers
The values of the totalizers are too high
See
Special Funct.\SYSTEM settings\Measuring\Quantity recall
. If this
menu item is activated, the values of the totalizer will be stored. The totalizer will continue
with this value at the start of the next measurement.
The values of the totalizers are too small
One of the totalizers has reached the upper limit and has to be reset to zero manually.
The sum of the totalizers is not correct
See
Special Funct.\SYSTEM settings\Measuring\Quant. wrapping
. The
sum of both totalizers (throughput) transmitted via the output is no longer valid after the
first overflow (wrapping) of one of the totalizers.