17 SuperUser mode
FLUXUS F401
17.5 Limit of the signal amplification
125
UMFLUXUS_F401V2-1EN, 2020-06-12
This display will only be indicated if
no
is selected in the menu item
Daily repeat
.
• Enter the starting time.
• Press ENTER.
This display will only be indicated if
setup
is selected.
• Select the measuring time.
• Press ENTER.
This display will only be indicated if
no
is selected in the menu item
Daily repeat
. The
max. measuring time is 12 h.
• Select the resting time.
• Press ENTER.
This display will only be indicated if
no
is selected in the menu item
Daily repeat
. The
max. resting time is 12 h.
17.5
Limit of the signal amplification
In order to prevent disturbing and/or pipe wall signals (e.g., if the pipe has run empty)
from being interpreted as useful signals, it is possible to define a max. signal amplifica
-
tion. If the signal amplification is greater than the max. signal amplification,
• the flow velocity will be marked as invalid. The physical quantity cannot be determined.
• a hash symbol (
#
) will be displayed after the unit of measurement (in case of a normal
error, (
?
) is displayed).
• Select
Special Funct.\SYSTEM settings\Measuring\Miscellaneous
.
• Press ENTER until the menu item
Gain threshold
is displayed.
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\
Gain threshold
Start of measure
02:00
Meas. duration ↕
1 hour
Sleep duration ↕
1 hour