9 Measurement
9.2 Measurement settings
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
60
Output range
When configuring an analog output, the output range has to be defined.
• Select
Special Funct.\SYSTEM settings\Proc. outputs\...\Output range I1
.
• Press ENTER.
• Select a list item.
– 4/20 mA
– other range...
• Press ENTER.
• If
other range
is selected, enter the values
Output MIN
and
Output MAX
.
• Press ENTER after each input.
Error output
In the following dialog, an error value can be defined which is to be transmitted if the
source item cannot be measured, e.g., if there are solids in the fluid.
• Select a list item for the error output, see Tab. 9.2.
• Press ENTER.
• If
Other value...
is selected, enter an error value. The value has to be within the
output range.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. outputs\...\Output range I1
Special Funct.\SYSTEM settings\Proc. outputs\...\Error-value I1
Notice!
The settings will be stored at the end of the dialog.
Summary of Contents for FLUXUS F401
Page 1: ...FLUXUS F401 UMFLUXUS_F401V2 1EN Operating instruction...
Page 158: ...Annex D Conformity declarations FLUXUS F401 2020 06 12 UMFLUXUS_F401V2 1EN 158...
Page 159: ......
Page 160: ......