9 Measurement
FLUXUS F401
9.2 Measurement settings
63
UMFLUXUS_F401V2-1EN, 2020-06-12
Terminal assignment
The terminals for the connection of the output are displayed.
• Press ENTER.
If the transmitter possesses a switchable current output, it is displayed whether it is ac
-
tive or passive (here: active).
Output function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the output.
• Enter a test value. It has to be within the output range.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions cor
-
rectly.
• Select
yes
to repeat the test,
no
to return to the menu item
SYSTEM settings
.
• Press ENTER.
Special Funct.\SYSTEM settings\Proc. outputs\...\Active loop I1
Special Funct.\SYSTEM settings\Proc. outputs\...\Output Test I1
Special Funct.\SYSTEM settings\Proc. outputs\...\I1= 10 mA\
Again?