9 Measurement
FLUXUS F401
9.1 Parameter input
55
UMFLUXUS_F401V2-1EN, 2020-06-12
9.1.2
Input of fluid parameters
Fluid temperature
At the beginning of the measurement, the fluid temperature is used for the interpolation
of the sound speed and therefore for the calculation of the recommended transducer dis
-
tance.
During the measurement, the fluid temperature is used for the interpolation of the density
and viscosity of the fluid.
The value entered here is used for the calculation if the fluid temperature is not mea
-
sured.
• Enter the fluid temperature. The value has to be within the operating temperature of the
transducers.
• Press ENTER.
9.1.3
Selection of the transducer
The transducer type has to be selected.
• Select
Standard
to use the standard transducer parameters stored in the transmitter.
• Press ENTER.
• Select the transducer frequency of the used transducer (see nameplate).
• Press ENTER.
Parameter\Medium Temperat.
Parameter\Transducer Type
Transducer
↕
CLAMPON 2MHZ