16 Advanced functions
16.1 Low-power mode
FLUXUS F401
2020-06-12, UMFLUXUS_F401V2-1EN
100
• Select a storage rate.
• Press ENTER.
This display will only be indicated if
Low-power mode
is activated.
16.1.3
Start of the measurement
• Select the program branch
Measuring
.
• Press ENTER until the menu item
Low-power mode
is displayed.
• Select
yes
to start the measurement in the low-power mode. Select
no
to start the
transmitter in normal measuring mode.
• Press ENTER.
The measurement is started, see Fig. 16.1.
For the setting of the measuring time see section 17.3.
Notice!
In order to start a measurement, the parameter input has to be completed.
Measuring\...\Low-power mode
Fig. 16.1: Measurement in the low-power mode
Storage Rate
↕
10 minutes
trans-
mitter
starts
measurement
resting phase
measuring phase
storage rate
measuring time
trans-
mitter
starts
measurement