Setting Up the Instrument in WDM Mode
66
FTB-5240S/5240BP
Setting Up WDM Analysis Parameters
You can set your pass/fail threshold limits in different ways depending on
the type of test you are performing.
Threshold Limit
Definition
None
No threshold limit is set. The results will be displayed without a
Pass/Fail verdict.
Min. only
The threshold limit is set for a minimum value only. The
Pass/Fail verdict is declared as Pass (in green), when the value
is equal to or greater than the minimum threshold set. The
verdict is declared as Fail (in red), when the value is below the
minimum threshold set.
Max. only
The threshold limit is set for a maximum value only. The
Pass/Fail verdict is declared as Pass (in green), when the value
is equal to or less than the maximum threshold set. The verdict
is declared as Fail (in red), when the value is above the
maximum threshold set.
Min. and Max.
The threshold limit is set for the minimum and maximum value.
The Pass/Fail verdict is declared as Pass (in green), when the
value is equal to or within the minimum and maximum
thresholds set. The Pass/Fail verdict is declared as Fail (in red),
when the value is beyond the minimum or maximum
thresholds set.
Use Default
This threshold limit option is available in the
Channel
Parameters
tab of
Channel Results
window only.
When this limit is set, the defaults set for the default channels in
the
Analysis Setup
tab will be applied to the channel.
Summary of Contents for FTB-5240BP
Page 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Page 8: ......
Page 14: ......
Page 110: ......
Page 190: ......
Page 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Page 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Page 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Page 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Page 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Page 424: ...A sample Drift txt report is shown below...
Page 428: ......
Page 434: ......
Page 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Page 444: ......
Page 446: ......
Page 530: ......