SCPI Command Reference
Optical Spectrum Analyzer
469
Product-Specific Commands—Description
:CALCulate[1..n]:RBW
Description
This command sets the RBW for OSNR. This
value is used in calculation only if RBW
Instrument is OFF (or 0).
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is 0.100 nm.
Syntax
:CALCulate[1..n]:RBW<wsp><RbwForOsnr>
Parameter(s)
RbwForOsnr:
The program data syntax for <RbwForOsnr> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
Selects the RBW for OSNR.
Range: [0.010 … 2.000] nm
Example(s)
CALC:RBW 0.1
See Also
CALCulate[1..n]:RBW?
CALCulate[1..n]:RBW:INSTrument
CALCulate[1..n]:RBW:INSTrument?
Summary of Contents for FTB-5240BP
Page 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Page 8: ......
Page 14: ......
Page 110: ......
Page 190: ......
Page 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Page 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Page 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Page 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Page 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Page 424: ...A sample Drift txt report is shown below...
Page 428: ......
Page 434: ......
Page 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Page 444: ......
Page 446: ......
Page 530: ......