Setting Up the Instrument in WDM Mode
Optical Spectrum Analyzer
69
Setting Up WDM Analysis Parameters
3.
Select the
Activate all thresholds
option, to manually set the global
threshold values. If this option is not selected, all the thresholds will be
deactivated, results are displayed without a Pass/Fail status and
Global
pass/fail status
will not be active in the
Global Results
tab.
4.
Enter values in the boxes as explained below:
³
Average signal power (dBm)
: indicates the sum of the signal
powers of all the peaks detected in the current acquisition, divided
by the total number of peaks.
³
Signal power flatness (dB)
: indicates the difference between the
maximum and minimum signal power values of the detected
peaks, in dB.
³
Average OSNR (dB)
: indicates the sum of the entire OSNR of the
peaks detected in the current acquisition, divided by the total
number of peaks.
³
OSNR flatness (dB)
: indicates the difference between the
maximum and minimum OSNR values of the detected peaks, in
dB.
Summary of Contents for FTB-5240BP
Page 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Page 8: ......
Page 14: ......
Page 110: ......
Page 190: ......
Page 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Page 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Page 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Page 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Page 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Page 424: ...A sample Drift txt report is shown below...
Page 428: ......
Page 434: ......
Page 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Page 444: ......
Page 446: ......
Page 530: ......