SCPI Command Reference
Optical Spectrum Analyzer
483
Product-Specific Commands—Description
:CALCulate[1..n]:WDM:SIGPower:
FLATness?
Description
This query returns the flatness signal power for a
channel.
*RST has no effect on the WDM measurement.
Syntax
:CALCulate[1..n]:WDM:SIGPower:FLATness?
Parameter(s)
None
Response Syntax
<PowerFlatness>
Response(s)
PowerFlatness:
The response data syntax for <PowerFlatness>
is defined as a <NR3 NUMERIC RESPONSE
DATA> element.
Flatness signal power from all channels (in dB).
Example(s)
INIT:IMM
CALC:WDM:SIGPower:FLATness?
See Also
CALCulate[1..n]:WDM:SIGPower:AVERage?
CALCulate[1..n]:WDM:SIGPower:INTegrated?
Summary of Contents for FTB-5240BP
Page 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Page 8: ......
Page 14: ......
Page 110: ......
Page 190: ......
Page 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Page 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Page 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Page 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Page 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Page 424: ...A sample Drift txt report is shown below...
Page 428: ......
Page 434: ......
Page 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Page 444: ......
Page 446: ......
Page 530: ......