Managing Results
Optical Spectrum Analyzer
315
Managing WDM Test Results
2.
Press
Global Parameters
.
3.
Select the
Global
Thresholds
tab.
4.
Select the
Activate all thresholds
option, to manually set the global
threshold values.
5.
Enter values in the boxes as explained below:
³
Average signal power (dBm)
: is the sum of all signal power of all
the peaks detected in the current acquisition, divided by the total
number of peaks.
³
Signal power flatness
(dB)
: is the difference between the
maximum and minimum signal power values of the detected
peaks, in dB.