Setting Up the Instrument in WDM Mode
64
FTB-5240S/5240BP
Setting Up WDM Analysis Parameters
³
Peak detection level (dBm)
: indicates the minimum power level
from where the peak can be considered as a signal.
³
RBW for OSNR (nm)
: indicates the resolution bandwidth for the
selected OSNR value. The resolution bandwidth of an OSA
determines its ability to deal with close optical channel spacing. It
is measured as the width of the response curve at half peak power
(i.e., 3 dB down) of the instrument to a monochromatic test signal.
The instrument’s RBW value is written below the graph on the right
(below wavelength offset).
³
Wavelength offset (nm)
: indicates the offset value applied on the
wavelength. You can use an offset to adjust your unit. This does not
replace a calibration performed at EXFO, but it can help you
achieve the specifications if you have determined that, for
example, your modules are used beyond the normal allowed use.
Entering a value in THz is not possible. The wavelength offset
appears below the graph, on the right, next to the
arrow.
³
Power offset (dB)
: indicates the offset value applied on the power.
You can use an offset to adjust your unit. This does not replace a
calibration performed at EXFO, but it can help you achieve the
specifications if you have determined that, for example, your
modules are used beyond the normal allowed use. The power
offset appears in the graph, on the left, above the
arrow.
To edit the power offset value in percentage, press
Edit %
button.
The percentage value entered in
Edit percentage
will be
converted to a corresponding equivalent value in dB.
³
Bandwidth at (dB)
: Set the power level used, relative to the
channel peak power, to compute the bandwidth.
Summary of Contents for FTB-5240BP
Page 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Page 8: ......
Page 14: ......
Page 110: ......
Page 190: ......
Page 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Page 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Page 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Page 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Page 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Page 424: ...A sample Drift txt report is shown below...
Page 428: ......
Page 434: ......
Page 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Page 444: ......
Page 446: ......
Page 530: ......