Setting Up the Instrument in Spectral Transmittance Mode
202
FTB-5240S/5240BP
Setting Up Spectral Transmittance Analysis Parameters
³
Channel distance (GHz or nm)
: indicates distance between the
channels. The value of channel distance will be set depending on
the selection made for the channel definition option. The channel
distance box will be enabled only when the channel center
wavelength option is set to custom.
³
Channel range (GHz or nm)
: indicates the limit inside which the
power values will be considered in the channel. Integrated power
is calculated on channel width.
³
Bandwidth 1 at (dB)
: Set the power level used, relative to the
channel peak power, to compute the bandwidth.
³
Bandwidth 2 at (dB)
: Set the power level used, relative to the
channel peak power, to compute the bandwidth.
³
Input wavelength offset (nm)
: indicates the offset value applied
on the input wavelength. You can use an offset to adjust your unit.
This does not replace a calibration performed at EXFO, but it can
help you achieve the specifications if you have determined that, for
example, your modules are used beyond the normal allowed use.
Entering a value in THz is not possible. The wavelength offset
appears below the graph, on the right, next to the
arrow.
³
Input power offset (dB)
: indicates the offset value applied on the
input power. You can use an offset to adjust your unit. This does
not replace a calibration performed at EXFO, but it can help you
achieve the specifications if you have determined that, for
example, your modules are used beyond the normal allowed use.
The power offset appears in the graph, on the left, above the
arrow.
Summary of Contents for FTB-5240BP
Page 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Page 8: ......
Page 14: ......
Page 110: ......
Page 190: ......
Page 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Page 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Page 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Page 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Page 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Page 424: ...A sample Drift txt report is shown below...
Page 428: ......
Page 434: ......
Page 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Page 444: ......
Page 446: ......
Page 530: ......