Setting Up the Instrument in EDFA Mode
Optical Spectrum Analyzer
225
Defining Preferences
2.
Select the
EDFA Results
tab.
3.
Select which parameters you want to display in the
Results
tab from
the list of available choices:
³
Name
: indicates the name of channel.
³
Center wavelength/frequency
: indicates the spectral
center-of-mass for the peak in that channel.
³
Input Signal Power
: indicates the signal power for the selected
channel (excludes noise).
³
Output Signal Power
: indicates the signal power for the selected
channel (excludes noise).
³
S %
: indicates the current output power according to the
measured output power (Output Signal Power / [Output Signal
Power + PASE]).
³
PASE
: indicates the power of the spontaneous emission amplified
by EDFA.
³
PSSE
: indicates the power of the spontaneous emission of the
source.
Summary of Contents for FTB-5240BP
Page 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Page 8: ......
Page 14: ......
Page 110: ......
Page 190: ......
Page 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Page 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Page 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Page 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Page 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Page 424: ...A sample Drift txt report is shown below...
Page 428: ......
Page 434: ......
Page 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Page 444: ......
Page 446: ......
Page 530: ......