SCPI Command Reference
498
FTB-5240S/5240BP
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:INBand:COUNt
Description
This command sets the number of In-Band scans
that will be averaged to produce final trace.
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is 64.
Syntax
:SENSe[1..n]:AVERage:INBand:COUNt<wsp>
<InbandCount>
Parameter(s)
InbandCount:
The program data syntax for <InbandCount> is
defined as a <numeric_value> element.
Selects the Number of scans used to inband for
the final scan.
Range: [4 ... 9999]
Example(s)
SENS:AVER:TYPE AVERAGING
SENS:AVER:INBand:COUN 4
SENS:AVER:COUN 12
INITiate:IMMediate
See Also
SENSe[1..n]:AVERage:INBand:COUNt?
SENSe[1..n]:AVERage:TYPE
Summary of Contents for FTB-5240BP
Page 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Page 8: ......
Page 14: ......
Page 110: ......
Page 190: ......
Page 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Page 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Page 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Page 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Page 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Page 424: ...A sample Drift txt report is shown below...
Page 428: ......
Page 434: ......
Page 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Page 444: ......
Page 446: ......
Page 530: ......