Setting Up the Instrument in Spectral Transmittance Mode
Optical Spectrum Analyzer
201
Setting Up Spectral Transmittance Analysis Parameters
3.
Under
Global analysis parameters
, define the following parameters
as needed:
³
Channel definition (GHz)
: indicates the limit inside which the
power values will be considered in the channel.
Centred on max peak
: Channel is centered on the lowest
insertion loss peak.
ITU Grid
: Select the nearest ITU channel from the peak with
lowest insertion loss.
CWDM
: Select the nearest CWDM channel from the peak with
lowest insertion loss.
Custom
: Channel is centered on value specified by the user.
³
Nominal wavelength or frequency (nm or Thz)
: indicates a
single value that represent the channels center wavelength (in nm)
or frequency (in THz). This field is editable only when
Channel
definition
is selected as
Custom
.
Summary of Contents for FTB-5240BP
Page 1: ...User Guide Optical Spectrum Analyzer for FTB 500 FTB 5240S 5240BP...
Page 8: ......
Page 14: ......
Page 110: ......
Page 190: ......
Page 282: ...Testing DWDM Systems 274 FTB 5240S 5240BP Starting a Measurement Drift Mode DFB Mode...
Page 391: ...Managing Results Optical Spectrum Analyzer 383 Managing EDFA Test Results...
Page 394: ...Managing Results 386 FTB 5240S 5240BP Managing EDFA Test Results...
Page 407: ...Managing Results Optical Spectrum Analyzer 399 Managing Trace Files 2 Press New...
Page 414: ...Managing Results 406 FTB 5240S 5240BP Managing Trace Files...
Page 424: ...A sample Drift txt report is shown below...
Page 428: ......
Page 434: ......
Page 437: ...Troubleshooting Optical Spectrum Analyzer 429 About 2 Press About...
Page 444: ......
Page 446: ......
Page 530: ......