22
WT.050.810.040.DE.IM.0519
Description
DEPOLOX
®
5 E flow cell
3.
3.5.2
3-electrode measuring cell
3.5.3
Sensors (optional)
pH combined measuring
and reference electrode
(W3T169297)
Design
Potentiostatic 3-electrode measuring cell
with platinum electrodes, tank with refer
-
ence electrolytes, two diaphragms, Ag/
AgCl drain system
Measurement range
0 to 50 mg/l (chlorine reference value)
Working temperature
range
0 to 50°C (32 to 122°F)
Area of application, pH
value
constant, within range pH 4 to pH 9
Influence of the pH
value
HOCl characteristic (for chlorine)
Measurable parame
-
ters
Free Cl
2
, ClO
2
, O
3
, KMnO
4
Conductivity
>= 100 µS/cm
Response time
t
90
= < 20 s
Typical output signal
20 µA/mg/l (for chlorine)
Storage temperature
-10 to +50°C (14 to 122°F)
Design
Combined measuring and reference elec
-
trode with universal membrane glass, salt
reserve, zirconium dioxide diaphragm,
polymerized solid electrolyte, Ag/AgCl
drain system
Measurement range
pH 0 to 12 (temporarily to pH 14)
Working temperature
range
-5 to +80°C (23 to 176°F)
Operating pressure
0 to 6 bar (6 × 10
5
Pa)
Minimum conductivity
of the sample water
50 µS/cm
Installation length
120 mm
Screw-in thread
PG 13.5
Storage temperature
-5 to +30°C (23 to 86°F)
Summary of Contents for WALLACE & TIERNAN DEPOLOX 5 E 700 P Series
Page 1: ...WALLACE TIERNAN DEPOLOX 5 E FLOW CELL INSTRUCTION MANUAL...
Page 56: ...56 WT 050 810 040 DE IM 0519 Maintenance DEPOLOX 5 E flow cell 5...
Page 72: ...72 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 76: ...76 WT 050 810 040 DE IM 0519 Declarations and certificates DEPOLOX 5 E flow cell 7...
Page 80: ...80 WT 050 810 040 DE IM 0519 Index DEPOLOX 5 E flow cell 8...