5. Installation
DEPOLOX® 400 M
48
EN
Fig. 16 Section, 3-electrode cell - color-coding
of the electrode connections on the
DEPOLOX® 5 C
A Working electrode (red cable)
B
Counter electrode (blue cable)
C
Reference electrode (white cable)
Fig. 17 Section, cap, DEPOLOX® 5 C flow cell
A Markings for opening and closing the cap
B
M12 threaded connection
5 Plug or screw the LED glow stick into the
location hole of the cell body cover.
6 Route the pH, ORP (Redox) or fluoride
sensor cable into the electronics module
through the corresponding cable gland.
7 Connect the sensor cable of the pH or flu-
oride sensor to the electronics module as
shown in the wiring diagram (chapter 9.).
5.13.2 Connect the sensor to the Y-style
flow-through assembly
8 Route the pH, ORP (Redox) or fluoride
sensor cable into the electronics module
through the corresponding cable gland.
9 Connect the sensor cable of the pH, ORP
(Redox) or fluoride sensor to the electro-
nics module as shown in the wiring dia-
gram (chapter 9.)
5.14 Electrical installation
A
B
C
A
B
DANGER
Risk of injury or death!
External voltages may still be connected
even if the operating voltage is switched off.
The electronics module must be de-energi-
zed before opening it.
WARNING
Risk of injury or damage to the device!
•
Only authorized and qualified electrici-
ans are permitted to install the DEPO-
LOX® 400 M and open the housing.
•
The electronics module may only be
wired in de-energized state.
•
Connect the electronics module in
accordance with the wiring diagram
(chapter 9.) and applicable local and
national regulations.
•
The electronics module may only be put
into operation when the housing is clo-
sed, and must be connected to protec-
tive earth.
•
Modifications to the device other than
those described in this operating
manual are not permissible.
•
High temperatures at the terminals of
the relays and the mains supply!
•
At high ambient temperatures, high
temperatures can occur at the termi-
nals, and the connected cables must be
designed to withstand such temperatu-
res.
NOTICE
•
The electronics module is not suitable
for electrical connection with perman-
ently installed cable conduits.
•
If the cable glands do not meet local ins-
tallation rules and regulations, these
glands must be replaced with suitable
ones.
•
The electronics module is equipped with
a flexible voltage supply input and
accepts AC voltages from 100 to 240
volts. Take the power consumption into
account when dimensioning (chapter
3.11.1).