DEPOLOX® 400 M
5. Installation
41
EN
5.5.4 Y-style flow-through assembly
1 Using mounting clamps (not included in
the scope of delivery), screw the Y-style
flow-through assembly to the solid wall
below the electronics module.
5.6
Removing and fittting the
housing cover
5.6.1 Flow cell
Removing
•
Remove the housing cover of the flow
cell. To do this, press both unlocking but-
tons (A) on the top of the housing and
carefully remove the cover toward the
front.
Fitting
•
Fit and engage the housing cover of the
flow cell. To do this, position the housing
cover at the bottom and carefully press it
upward until the housing cover engages
on the unlocking buttons.
Fig. 1
View of top of housing cover
A Unlocking buttons
5.6.2 Electronics module
Removing
1 Release the four screws on the housing
cover.
2 Remove the housing cover carefully.
3 Hook the housing cover into the holders
(A) on the basic housing.
Fig. 2
Section, housing cover
A Holders
Fitting
1 Place the housing cover carefully on the
basic housing.
2 Tighten the housing screws by hand (to a
maximum torque of 0.5 ft-lb ± 0.11 ft-lb
(0.7 Nm ± 0.15 Nm)).
A
A
A