6. Operation
DEPOLOX® 400 M
54
EN
Chlorine measurement parameters
pH measurement parameters
Measurement
Range
Range end:
1.00 / 2.00 / 5.00 / 10.00 / 20.00 mg/l or
ppm
200 / 500 µg/l or ppd
Unit:
µg/l / mg/l / ppd / ppm
Sensor type:
O3sel(M) / ClO2sel(M) / Cl-tot(M) / Cl2
free(M) / KMnO4 / O3 / ClO2 /Cl2 free,
Cl2++
Measurement filter:
off/low/middle/strong
Upot:
-1000 to 1000
(factory setting 250 mV)
µA-Range:
10 µA / 100 µA / 1000 µA
(factory setting 100 µA)
Notice: Depending on the sensor and the
size of the measurement, the µA-Range may
need to be adjusted. If the µA-Range is not
sufficient, select the next higher µA-Range.
Limits
Limits value I
Max:
within range
Min:
within range
Hysteresis:
0.01 to 0.25
Limits value II
Max:
within range
Min:
within range
Hysteresis:
0.01 to 0.25
Maintenance intervals
Calibration
Message:
On/Off
Interval:
2 to 90
Next maintenance on:
Date
Reset maintenance:
Execute
Cell
Message:
On/Off
Interval:
10 to 365
Next maintenance on:
Date
Reset maintenance:
Execute
Measurement
Range
Range end:
pH 0.00 to pH 14.00
Range start:
pH 0.00 to pH 14.00
Measurement filter:
off/low/middle/strong
Limits
Limits value I
Max:
within range
Min:
within range
Hysteresis:
1 to 25
Limits value II
Max:
within range
Min:
within range
Hysteresis:
1 to 25
Maintenance intervals
Calibration
Message:
On/Off
Interval:
2 to 90
Next maintenance on:
Date
Reset maintenance:
Execute
Cell
Message:
On/Off
Interval:
10 to 365
Next maintenance on:
Date
Reset maintenance:
Execute