5. Installation
DEPOLOX® 400 M
50
EN
5.15 Initial commissioning
Following complete mechanical and electrical
installation, initial commissioning can be car-
ried out. Please check that the following con-
ditions are met:
•
Electronics module, flow cell or Y-style
flow-through assembly are installed.
•
Housing covers are fitted.
•
Optional strainer is installed.
•
Sample water inlet and outlet are connec-
ted.
•
Sensors are installed in the flow cell or Y-
style flow-through assembly.
•
The sensors are connected to the electro-
nics module.
•
The electronics module is wired in accor-
dance with the wiring diagram (circuit dia-
gram) and local regulations.
•
Ensure that all transport protection was
removed.
•
Check all connections for leakage.
•
Sample water monitoring is active (chap-
ter 3.6.6).
1 Switch on power supply.
2 Perform initial configuration:
•
Select the language
•
Set the date and time
•
Enter the system name
3
Chlorine, chlorine dioxide, ozone or
potassium permanganate measurement
(DES sensors)
•
Check the limit values 1 and 2 (“Min"
and “Max"), adjust if necessary.
•
Check the measurement range, adjust
if necessary.
4 pH, OPR (Redox) and fluoride measure-
ment
•
Check the limit values 1 and 2 for pH/
ORP (Redox)/fluoride (Min and Max),
adjust if necessary.
•
Check the measurement range for pH/
OPR (Redox)/fluoride, adjust if neces-
sary.
5 Configure mA outputs.
6 Carry out initial calibration of the sensors
used.
DANGER
Risk of injury or death!
The DEPOLOX® 400 M must not be operated
with flammable liquids.
ATTENTION
Risk of injury or damage to the device!
•
To ensure safe and correct commissio-
ning, knowledge of the operation, con-
nected electrical load, measurement
signals, cable assignment and fuse pro-
tection of the connected devices and
machines and the relevant safety regu-
lations is required.
•
Startup of the DEPOLOX® 400 M may
therefore only be performed by quali-
fied and authorized electricians.
•
Incorrectly connected devices can be
damaged, possibly irreparably, or cause
faults in other equipment when they are
switched on or in operation.
•
Ensure that the measuring and control
cables are not confused or make contact
with one another.
•
Never connect or disconnect any cables
to which voltage is applied.
WARNING
Risk of injury or damage to the device!
•
When connecting the DEPOLOX® 400 M
to the power supply, a 6A back-up fuse
must be used in the mains supply line.