DEPOLOX® 400 M
3. Description
25
EN
3.11.5 Sensors
pH sensor
Version
Single-rod measuring cell with universal membrane glass, salt reserve,
zirconium dioxide diaphragm, polymerized solid electrolyte, Ag/AgCl
reference electrode
Range
pH 0 to 12 (temporarily to pH 14)
Working tempera-
ture range
23 to 176°F (-5 to +80°C)
Operating pressure
6 x 10
5
Pa (0 to 6 bar)
Minimum conducti-
vity of the sample
water
50 µS/cm
Installation length
4.72“ (120 mm)
Screw-in thread
PG 13.5
Storage temperature 23 to 86°F (-5 to +30°C)
ORP sensor (Redox)
Version
Combination electrode with platinum electrode, salt reserve,
zirconium dioxide diaphragm, polymerized solid electrolyte, Ag/AgCl
reference electrode
Range
±2000 mV
Working tempera-
ture range
23 to 176°F (-5 to +80°C)
Operating pressure
0 to 6 bar (6 x 10
5
Pa)
Minimum conducti-
vity of the sample
water
50 µS/cm
Installation length
120 mm
Screw-in thread
PG 13.5
Storage temperature
23 to 86°F (-5 to +30°C)
Fluoride sensor
Sensor input
Fluoride single-rod electrode
Range
0.2 to 20.0 mg/l
pH range