7. Maintenance
DEPOLOX® 400 M
74
EN
7.
Maintenance
7.1 Automatic maintenance and
calibration check intervals
The electronics module offers the option of
using automatic maintenance and calibration
check intervals for disinfection, and pH, ORP
(Redox) and fluoride measurement. In as-deli-
vered status, the maintenance and calibration
intervals are disabled. To enable the mainte-
nance intervals for disinfection, pH and fluo-
ride measurement:
1 Call up the Home menu.
2 Select the desired measurement (e.g.
Chlorine free). The “Measurement“ menu
opens.
3 Tap the
Settings menu. The settings
menu for the selected measurement
opens.
4 “Maintenance interval” menu opens.
5 To enable the maintenance interval, set
the parameter “Message” to “On” under
the heading Cells and, under the parame-
ter “Interval,” enter the number of days
after which you wish to receive a remin-
der to carry out sensor calibration.
When the set interval expires, an orange
message appears on the electronics
module reminding you to carry out sensor
calibration. The message can be acknow-
ledged via the message system.
To reset the maintenance interval, tap the
“Execute” button under the heading
“Cell.” The date on which the next mainte-
nance is due is recalculated and displayed
under “next maintenance on.”
7.2 Maintenance intervals
regularly
•
Sample water monitoring (chapter 3.6.6)
•
DEPOLOX® 5 C: Check electrolyte level
daily
•
Check the flow cell, including all screw
connections, for leakage
weekly
•
DEPOLOX® 5 C: Check the electrode clea-
ning sand
every six months
•
DEPOLOX® 5 C: Replace electrolyte level
•
DEPOLOX® 5 C: Replace the electrode
cleaning sand
annually
•
DEPOLOX® 5 C: Check diaphramgs
2 years
•
DEPOLOX® 5 C: Replacing reference
electrode
5 years
•
Replace battery of the electronics module
•
Replace battery of the Impedance conver-
ter
as required
•
VariaSens C: Cleaning or replacing the fine
filter
•
DEPOLOX® 5 C: Check diaphramgs (depen-
ding on sample water quality)
•
DEPOLOX® 5 C: Replacing reference elec-
trode
•
Clean or replace optional strainer
in accordance with standard or local
regulations
•
Comparative measurement of chlorine,
pH, ORP (Redox) and fluoride. Carry out
calibration.
DANGER
Risk of injury or death!
External voltages may still be connected
even if the operating voltage is switched off.
NOTICE
•
Liability for defects can only be accepted
if maintenance work is performed as
specified.
•
Adhere to the applicable standards and
national and regional regulations.
•
Observe maintenance intervals of the
sensors. See instruction manual of sen-
sors!