DEPOLOX® 400 M
5. Installation
51
EN
5.16 Replace 4-way mA output card
1 Disconnect the electronics module from
the power supply.
2 Remove the housing cover of the electro-
nics module.
3 Install mA output card at position A and
make sure that the holders engage. Note
the correct installation direction.
4 Plug in terminal block at position B.
5 Wire in accordance with the wiring dia-
gram (chapter 9.).
6 Fit the housing cover.
7 Connect the power supply again.
8 The electronics module automatically
detects the installed modules and enables
the corresponding settings menus.
9 Configure mA outputs as required.
10 Check all menu settings and configure if
necessary.
Fig. 19 View, installation of the 4-way mA
output card
5.17 Shut-down
1 Disconnect
the
DEPOLOX® 400 M from
the power supply.
2 Drain the sample water supply line and
drainage line.
3 Remove the housing cover of the flow
cell.
4 Drain the cell body via the flow cell drain
valve.
5 Flush out the cleaning sand.
6 Dismantle the filter unit and the check
valve housing (chapter 7.7).
7 When the remaining water has drained
from the flow control valve, refit the filter
housing and the check valve housing.
8 Remove the sensors from the location
hole in the cover of the cell body and dis-
connect from the electronics module.
9 Shut down the sensors. See the relevant
operating instructions for the sensors.
•
Plug in the pH or OPR (Redox) sensor
in the transport container with dilu-
ted electrolytes (KCl tank with stand
and 5 ml KCl solution).
•
Store the sensor in a frost-protected
location.
10 Replace and engage the housing cover.
5.18 Start-up
See chapter 5.15.
DANGER
Risk of injury or death!
External voltages may still be connected
even if the operating voltage is switched off.
NOTICE
After changes the electronics module resets
all parameter to factory settings.
A
B
DANGER
Risk of injury or death!
External voltages may still be connected
even if the operating voltage is switched off.