The
Reagents
tab shows which reagents must be loaded. In the left part of the
window there is a graphic display of the rack for the secondary reagents and its
position for the test run to be performed. The reagents used are shown in
different colours and are assigned to a position in the rack. The legend to the
right describes the assignment of the colours to the tests.
Ensure that there is enough liquid in the bottles.
Figure 7-54: ELISA reagents
After the conjugate step the pipetting unit needs to be rinsed with
sulfuric acid (0.5 mol/l). This dilution needs to be inserted in the device
by the user (Sprinter XL Position D05). The software does not request
this dilution.
The
Time Schedule
tab shows the detailed temporal progression of the
worklist. Here you can see the status (“Running" or “Done”) and the scheduling
of the individual processing steps of the different tests.
The times displayed were determined theoretically. They are compared with
real times during processing and are recalculated, if necessary, at the end of a
step.