- 28 -
7 Operating instructions
M e d i z i n i s c h e
L a b o r d i a g n o s t i k a
A G
EUROIMMUN
Before starting a worklist, the system liquid must be flushed in to remove air
bubbles from the tube. You can flush in the liquid using the service menu.
To reach the service menu, press the
Service
button. The service
menu will offer you a list with different service points of the Sprinter
XL.
Figure 7-5:
Service Menu1
Select the
Maintenance – Prime Procedure
menu item under
Service
operations
and press
Start
. In the
Description
field, you will get
instructions which you must follow to perform this step.
Figure 7-6:
Service Menu Description1
Ensure that the system liquid container is sufficiently filled. When flushing
in liquid, watch out for any bubbles which may appear in the system tube. If
necessary, repeat the process.
To start the service, press the
Continue
button. The Sprinter XL
flushes the needle washing station with system liquid.