9 Cleaning and maintenance
M e d i z i n i s c h e
L a b o r d i a g n o s t i k a
A G
EUROIMMUN
Remove the system filter.
Remove the metal plate.
Lift out the insert and clean it with water, a brush and
a mild detergent.
Thoroughly rinse with distilled water.
Remove the end of the system tube from the station
and insert it in a vessel with 70% alcohol.
Then repeat the procedure twice with distilled water.
Perform the
Prime Procedure
in the Sprinter XL
software under the
Service
menu item.
Then rinse thoroughly with distilled water (5x).
Decontamination
of the system
filter