- 36 -
7 Operating instructions
M e d i z i n i s c h e
L a b o r d i a g n o s t i k a
A G
EUROIMMUN
Figure 7-19:
Sample Selection
To de-select samples, hold down the control key (Control on the keyboard) and
left-click into the field to be de-selected. It turns white again.
To delete samples, click on the position number in the tube column
so that the line selected turns blue then click on the trash can
symbol on the left margin.
Click on the test name in the header of a column displayed in yellow to confirm
the test for the next worklist. Following activation, the colour of the column
changes from yellow to green and the test name changes from black to red.
Only tests which are confirmed in this way are taken into account in the next
test run.
Figure 7-20:
Test confirmation
One dilution, the initial dilution, is performed by default for each selected
sample. It depends on the programming which initial dilution corresponds to
which respective test.
If additional dilutions for IFA should be performed for titer determination, left-
click twice on the “T” symbol above the test name in the corresponding test
column. A window with the titration dilutions opens.
De-selecting
samples
Deleting
samples
Confirming
a test
Additional
dilutions