Settings of the ETL Interface
Instruction Manual ATS 400 with SC-NG
Version 4.03
54
10.6
Using the Buzzer
The built in buzzer will be used if the checkbox
Beep I.O.
resp.
Beep N.I.O.
is
active. The duration of the tone depends on the time given in the field
BeepTime
.
The buzzer will also be activated when powering on the system if the checkbox for
Beep I.O.
is active.
10.7
Disable SC, Con
In the case this checkbox is inactive the signals
Security circuit closed
and
Contacting closed
are routed to the ETL Interface.
Is this checkbox active this signals can be used to control an external equipment. In
ETL DataView
3 this will be configured in the test steps on the property page
Matrix
, using your own application this controlled by the parameter
Contacting
.
10.8
KeepPolarity
Is this checkbox inactive an optional relais matrix will be switched to the default
position at the end of a test plan.
Is this checkbox active the position of an optional relais matrix will be kept active
after the end of a test plan.
10.9
Hold matrix at start with contacting
Uses a test step a matrix the matrix will be turned off after 2 minutes when the step
will not bestarted.
In this situation you cannot start with contact moitoring.
Setting this check box and contact monitoring is a strt condition the matrix will not
be turned off after 2 minutes.
10.10
Warn lamp flashing
Has the device IO-CPU firmware 33244 or newerand safety circuit firmware 2 or
newer this checkbox will be displayed.
Is this checkbox active the red warning lamp will be flashing during an active test.
10.11
Adaptive HV timing
Has the device IO-CPU firmware 33284 or newer this checkbox will be displyed.
Is this checkbox active the start of the test will be modified during a high voltage
test with AC current.