Information for safe Operation
Instruction Manual ATS 400 with SC-NG
Version 4.03
21
5
Information for safe Operation
5.1
General Safety Information
DANGER
Depending on the type, the test system can supply high potentials with high
efficiency. The safety measures demanded in VDE 0104 (En 50 191) are to observe.
The proper status of the main cable and the test object connection cables is to
observe via visual every day before starting the testing work.
Damaged parts must be replaced or put out of operation.
No start-up when there are obvious defects!
Only workshops who have been authorised be ETL are allowed to open the device
and overhaul it! There are no parts in the interior of the device that can be
replaced by the operator.
The ATS 400 is a device of protection class I.
The protective earth conductor connection of the used main cable and the power
socket must be proper. Every interruption of the protective earth conductor can
lead to the fact that the device might become dangerous. Therefore an interruption
of the protective earth conductor is illegal.
5.2
Special Safety Information for High Voltage Testing and other
dangerous Kinds of Testing
5.2.1
Testing with High Voltage Test Pistols
5.2.1.1
Fusing the Work Place
When using two high voltage test pistols, the test station must be organised
according to EN 50191 (DIN VDE 0104) (Electric Test Stations), paragraph for
„Test Stations without obligatory protection against contact“:
5.2.1.2
Protecting Outsiders
DANGER
Outsiders must be protected from accidently touching the test object (and so from
touching the high voltage) by:
Cordoning off the test area
Warning signs WS1 and ZS 1 “High voltage, danger of life”
Warning lamps, red-green combination
Disctances to the high voltage according to EN 50191
Instructions
5.2.1.3
Protecting the Test Person
The test person is protected by: